.5SU
/ .5SUM
![[Sensor .5SU / .5SUM]](images/sensr-dot5SU5SUM.GIF)
Dimensions
shown in
inches (mm).
back
to top
|
|
Sensor
Type
|
.5SU
|
.5SUM
|
| Target
material |
non-mag
magnetic |
Yes
No |
No
Yes |
Measuring
range
(+10% nominal offset) |
mil*
mm |
20
0.5 |
20
0.5 |
| Offset |
mil*
mm |
2
0.05 |
2
0.05 |
Nonlinearity
(±1% of measuring range) |
±mil*
±mm |
0.2
0.005 |
0.2
0.005 |
Static
Resolution - mid range
Analog |
mil*
µm |
0.004
0.1 |
0.004
0.1 |
Sensitivity
<= mid range
Customer adjustable |
mV/mil*
mV/mm |
50
1000 |
50
1000 |
Output,
full range
Voltage analog
Factory setting
Customer adjustable |
Vdc(eng.)
Vdc(met.) |
1.000
0.500 |
1.000
0.500 |
Output,
full range
Current
Impedance |
mA
ohms |
Maximun of 15
Less than 5 |
Thermal
sensitivity
shift <= mid range
(Sensor/target interface) |
0.1%
FSO/°F without temp.comp.;
0.02% FSO/°F with
temp. comp. |
| **Frequency
response |
Static
to 50 kHz (-3dB pt.) |
*mil = 0.001 inch
**Maximum frequency
response is sensitivity dependent.
At maximum sensitivity, the -3 dB level is 10 kHz.
At 10% of maximum sensitivity, frequency response is 50 kHz (optimum).
|
1S
/ 1SM
![[Sensor 1S / 1SM]](images/sensr-1S1SM.GIF)
Dimensions
shown in
inches (mm).
Metric threads
optional
(M7 x 1).
back
to top
|
|
Sensor
Type
|
1S
|
1SM
|
| Target
material |
non-mag
magnetic |
Yes
No |
No
Yes |
Measuring
range
(+10% nominal offset) |
mil*
mm |
40
1.0 |
40
1.0 |
| Offset |
mil*
mm |
2
0.05 |
2
0.05 |
Nonlinearity
(±1% of measuring range) |
±mil*
±mm |
0.4
0.01 |
0.4
0.01 |
Static
Resolution - mid range
Analog |
mil*
µm |
0.004
0.1 |
0.004
0.1 |
Sensitivity
<= mid range
Customer adjustable |
mV/mil*
mV/mm |
20
1000 |
20
1000 |
Output,
full range
Voltage analog
Factory setting
Customer adjustable |
Vdc(eng.)
Vdc(met.) |
0.800
1.000 |
0.800
1.000 |
Output,
full range
Current
Impedance |
mA
ohms |
Maximun of 15
Less than 5 |
Thermal
sensitivity
shift <= mid range
(Sensor/target interface) |
0.1%
FSO/°F without temp.comp.;
0.02% FSO/°F with
temp. comp. |
| **Frequency
response |
Static
to 50 kHz (-3dB pt.) |
*mil = 0.001 inch
**Maximum frequency
response is sensitivity dependent.
At maximum sensitivity, the -3 dB level is 10 kHz.
At 10% of maximum sensitivity, frequency response is 50 kHz (optimum).
|
1SU
/ 1SUM
![[Sensor 1SU / 1SUM]](images/sensr-1SU1SUM.GIF)
Dimensions
shown in
inches (mm).
Metric threads
optional
(M7 x 1).
back
to top
|
|
Sensor
Type
|
1SU
|
1SUM
|
| Target
material |
non-mag
magnetic |
Yes
No |
No
Yes |
Measuring
range
(+10% nominal offset) |
mil*
mm |
50
1.25 |
50
1.25 |
| Offset |
mil*
mm |
2
0.05 |
2
0.05 |
Nonlinearity
(±1% of measuring range) |
±mil*
±mm |
0.5
0.0125 |
0.5
0.0125 |
Static
Resolution - mid range
Analog |
mil*
µm |
0.005
0.1 |
0.005
0.1 |
Sensitivity
<= mid range
Customer adjustable |
mV/mil*
mV/mm |
20
1000 |
20
1000 |
Output,
full range
Voltage analog
Factory setting
Customer adjustable |
Vdc(eng.)
Vdc(met.) |
1.000
1.250 |
1.000
1.250 |
Output,
full range
Current
Impedance |
mA
ohms |
Maximun of 15
Less than 5 |
Thermal
sensitivity
shift <= mid range
(Sensor/target interface) |
0.1%
FSO/°F without temp.comp.;
0.02% FSO/°F with
temp. comp. |
| **Frequency
response |
Static
to 50 kHz (-3dB pt.) |
*mil = 0.001 inch
**Maximum frequency
response is sensitivity dependent.
At maximum sensitivity, the -3 dB level is 10 kHz.
At 10% of maximum sensitivity, frequency response is 50 kHz (optimum).
|
2S
![[Sensor 2S]](images/sensr-2S.GIF)
Dimensions
shown in
inches (mm).
Metric threads
optional
(M12 x 1).
back
to top
|
| Sensor
Type |
2S |
| Target
material |
non-mag
magnetic |
Yes
Yes |
Measuring
range
(+10% nominal offset) |
mil*
mm |
100
2.5 |
| Offset |
mil*
mm |
10
0.25 |
Nonlinearity
(±1% of measuring range) |
±mil*
±mm |
1.0
0.025 |
Static
Resolution - mid range
Analog |
mil*
µm |
0.01
0.3 |
Sensitivity
<= mid range
Customer adjustable |
mV/mil*
mV/mm |
10
1000 |
Output,
full range
Voltage analog
Factory setting
Customer adjustable |
Vdc(eng.)
Vdc(met.) |
1.000
2.500 |
Output,
full range
Current
Impedance |
mA
ohms |
Maximun of 15
Less than 5 |
Thermal
sensitivity
shift <= mid range
(Sensor/target interface) |
0.1%
FSO/°F without temp.comp.;
0.02% FSO/°F with
temp. comp. |
| **Frequency
response |
Static
to 50 kHz (-3dB pt.) |
*mil = 0.001
inch
**Maximum frequency
response is sensitivity dependent.
At maximum sensitivity, the -3 dB level is 10 kHz.
At 10% of maximum sensitivity, frequency response is 50 kHz (optimum).
|
6C
![[Sensor 6C]](images/sensr-6C.GIF)
Dimensions
shown in
inches (mm).
back
to top
|
| Sensor
Type |
6C
|
| Target
material |
non-mag
magnetic |
Yes
Yes |
Measuring
range
(+10% nominal offset) |
mil*
mm |
250
6.4 |
| Offset |
mil*
mm |
30
0.75 |
Nonlinearity
(±1% of measuring range) |
±mil*
±mm |
2.5
0.064 |
Static
Resolution - mid range
Analog |
mil*
µm |
0.025
0.6 |
Sensitivity
<= mid range
Customer adjustable |
mV/mil*
mV/mm |
5
100 |
Output,
full range
Voltage analog
Factory setting
Customer adjustable |
Vdc(eng.)
Vdc(met.) |
1.250
0.640 |
Output,
full range
Current
Impedance |
mA
ohms |
Maximun of 15
Less than 5 |
Thermal
sensitivity
shift <= mid range
(Sensor/target interface) |
0.1%
FSO/°F without temp.comp.;
0.02% FSO/°F with
temp. comp. |
| **Frequency
response |
Static
to 50 kHz (-3dB pt.) |
*mil = 0.001
inch
**Maximum frequency
response is sensitivity dependent.
At maximum sensitivity, the -3 dB level is 10 kHz.
At 10% of maximum sensitivity, frequency response is 50 kHz (optimum).
|
8C
![[Sensor 8C]](images/sensr-8C.GIF)
Dimensions
shown in
inches (mm).
back
to top
|
| Sensor
Type |
8C
|
| Target
material |
non-mag
magnetic |
Yes
Yes |
Measuring
range
(+10% nominal offset) |
mil*
mm |
500
12.7 |
| Offset |
mil*
mm |
50
1.25 |
Nonlinearity
(±1% of measuring range) |
±mil*
±mm |
5.0
0.127 |
Static
Resolution - mid range
Analog |
mil*
µm |
0.05
1.3 |
Sensitivity
<= mid range
Customer adjustable |
mV/mil*
mV/mm |
2
100 |
Output,
full range
Voltage analog
Factory setting
Customer adjustable |
Vdc(eng.)
Vdc(met.) |
1.000
1.270 |
Output,
full range
Current
Impedance |
mA
ohms |
Maximun of 15
Less than 5 |
Thermal
sensitivity
shift <= mid range
(Sensor/target interface) |
0.1%
FSO/°F without temp.comp.;
0.02% FSO/°F with
temp. comp. |
| **Frequency
response |
Static
to 50 kHz (-3dB pt.) |
*mil = 0.001
inch
**Maximum frequency
response is sensitivity dependent.
At maximum sensitivity, the -3 dB level is 10 kHz.
At 10% of maximum sensitivity, frequency response is 50 kHz (optimum).
|
10C
![[Sensor 10C]](images/sensr-10C.GIF)
Dimensions
shown in
inches (mm).
back
to top
|
| Sensor
Type |
10C
|
| Target
material |
non-mag
magnetic |
Yes
Yes |
Measuring
range
(+10% nominal offset) |
mil*
mm |
750
19.1 |
| Offset |
mil*
mm |
100
2.5 |
Nonlinearity
(±1% of measuring range) |
±mil*
±mm |
7.5
0.191 |
Static
Resolution - mid range
Analog |
mil*
µm |
0.075
1.9 |
Sensitivity
<= mid range
Customer adjustable |
mV/mil*
mV/mm |
1
100 |
Output,
full range
Voltage analog
Factory setting
Customer adjustable |
Vdc(eng.)
Vdc(met.) |
0.750
1.910 |
Output,
full range
Current
Impedance |
mA
ohms |
Maximun of 15
Less than 5 |
Thermal
sensitivity
shift <= mid range
(Sensor/target interface) |
0.1%
FSO/°F without temp.comp.;
0.02% FSO/°F with
temp. comp. |
| **Frequency
response |
Static
to 50 kHz (-3dB pt.) |
*mil = 0.001
inch
**Maximum frequency
response is sensitivity dependent.
At maximum sensitivity, the -3 dB level is 10 kHz.
At 10% of maximum sensitivity, frequency response is 50 kHz (optimum).
|
10CU
![[Sensor 10CU]](images/sensr-10CU.GIF)
Dimensions
shown in
inches (mm).
back
to top
|
| Sensor
Type |
10CU
|
| Target
material |
non-mag
magnetic |
Yes
Yes |
Measuring
range
(+10% nominal offset) |
mil*
mm |
1000
25.4 |
| Offset |
mil*
mm |
100
2.5 |
Nonlinearity
(±1% of measuring range) |
±mil*
±mm |
10.0
0.254 |
Static
Resolution - mid range
Analog |
mil*
µm |
0.1
2.5 |
Sensitivity
<= mid range
Customer adjustable |
mV/mil*
mV/mm |
1
100 |
Output,
full range
Voltage analog
Factory setting
Customer adjustable |
Vdc(eng.)
Vdc(met.) |
1.000
2.540 |
Output,
full range
Current
Impedance |
mA
ohms |
Maximun of 15
Less than 5 |
Thermal
sensitivity
shift <= mid range
(Sensor/target interface) |
0.1%
FSO/°F without temp.comp.;
0.02% FSO/°F with
temp. comp. |
| **Frequency
response |
Static
to 50 kHz (-3dB pt.) |
*mil = 0.001
inch
**Maximum frequency
response is sensitivity dependent.
At maximum sensitivity, the -3 dB level is 10 kHz.
At 10% of maximum sensitivity, frequency response is 50 kHz (optimum).
|
12CU
![[Sensor 12CU]](images/sensr-12CU.GIF)
Dimensions shown
in
inches (mm).
back
to top
|
| Sensor
Type |
12CU
|
| Target
material |
non-mag
magnetic |
Yes
Yes |
Measuring
range
(+10% nominal offset) |
mil*
mm |
2000
50.8 |
| Offset |
mil*
mm |
200
5.0 |
Nonlinearity
(±1% of measuring range) |
±mil*
±mm |
20.0
0.508 |
Static
Resolution - mid range
Analog |
mil*
µm |
0.2
5.1 |
Sensitivity
<= mid range
Customer adjustable |
mV/mil*
mV/mm |
0.5
100 |
Output,
full range
Voltage analog
Factory setting
Customer adjustable |
Vdc(eng.)
Vdc(met.) |
1.000
5.080 |
Output,
full range
Current
Impedance |
mA
ohms |
Maximun of 15
Less than 5 |
Thermal
sensitivity
shift <= mid range
(Sensor/target interface) |
0.1%
FSO/°F without temp.comp.;
0.02% FSO/°F with
temp. comp. |
| **Frequency
response |
Static
to 50 kHz (-3dB pt.) |
*mil = 0.001
inch
**Maximum frequency
response is sensitivity dependent.
At maximum sensitivity, the -3 dB level is 10 kHz.
At 10% of maximum sensitivity, frequency response is 50 kHz (optimum).
|
|